Journal of Neural Engineering· Sulagna Sahu, Munish Chauhan, Saurav Z K Sajib, Enock Boakye and Rosalind J Sadleir·· 6 天前AI 评分34

计算建模揭示 tACS 对三叉神经分支的非预期激活阈值

Mapping unintended trigeminal nerve activation during transcranial alternating current stimulation (tACS): a computational study

AI 导读

一项计算建模研究用 MRI 有限元头部模型结合 NEURON 有髓感觉轴突模型,量化了临床 tACS 参数下 7 条三叉神经分支的激活阈值。在 F3–F4 与 Fp1–Fp2 两种电极布置下,40 Hz 及以上时部分分支阈值低于 2 mA,且 60 Hz 正弦波阈值始终低于 60 Hz 双相脉冲。

正文 · 原文

Objective. To systematically characterize activation thresholds for peripheral trigeminal nerve branches under clinical transcranial alternating current stimulation (tACS) parameters using a computational modeling approach. Approach. An MRI-derived finite element head model was coupled with multicompartment NEURON cable models of myelinated sensory axons (4–14 µm diameter). Seven trigeminal branches were manually delineated. Activation thresholds were computed across eight sinusoidal frequencies (5–100 Hz) for two electrode montages (F3–F4 and Fp1–Fp2). A 60 Hz biphasic pulse enabled direct comparison with established transcutaneous nerve stimulation devices. Sensitivity analyses assessed robustness to nerve orientation and depth, and electrode displacement. Main Results. Threshold decreased non-linearly with increasing frequency, plateauing above 40 Hz. Under F3–F4, the auriculotemporal and zygomaticotemporal nerves, and under Fp1–Fp2, the supraorbital and supratrochlear nerves fell below 2 mA threshold at ⩾40 Hz. Thresholds were largely independent of axon diameter above 40 Hz. Thresholds for 60 Hz sinusoidal waveforms were consistently lower than 60 Hz biphasic pulse condition. Sensitivity analyses confirmed sub-2 mA thresholds were preserved for tested perturbations. Significance. Select trigeminal branches are activated below the standard 2 mA clinical dose across both montages. This study provides preliminary indication that peripheral trigeminal recruitment is a predictable, montage-dependent, and mechanistically plausible contributor to tACS effects that warrants explicit consideration in trial design and result interpretation.

来源:Journal of Neural Engineering · iopscience.iop.org

原文链接:https://iopscience.iop.org/article/10.1088/1741-2552/aeb074